Canberra MIP 10 Digital Now Mirion Technologies
desktop ratemeter / survey meter with scaler-timer, for smart (CSP) and Nardeux probes
- Detector
- none built in; two probe inputs, one for CSP smart probes and one for older Nardeux probes, so two probes can be connected at once (e.g. SG 1R, SAB 100, SB 100, SPAB 15 and SBM series CSP probes)
- Detects
- alpha, beta, gamma and X-ray, depending on the probe connected
- Range
- units shown depend on the probe: c/s, Bqeq, Bqeq/cm2, Sv/h, Sveq/h (Nardeux probes via MIP10/MIP21 adaptor show c/s only); semi-log bargraph over four decades, as fast as 1/4 s; digital readout 1 s; scaler/timer mode 1 to 1000 s integration
Bench or pedestal ratemeter for contamination and dose rate probes at control points and in labs, with per-probe alarm setpoints.
Successor to the MIP 10; same dimensions so it fits MIP 10 accessories; 62 x 213 x 225 mm, 1800 g (4 lb); metal housing; IP 40; -10 to +50 C; built-in NiMH backup batteries, over 40 h with CSP probes and over 15 h with Nardeux probes, full charge under 5 h; mains 100-240 V ac 50/60 Hz; CSP alarm setpoints (10 values) are stored in the probe, Nardeux setpoint in the meter; optional 1 m pedestal; Avior is the single-input CSP-only alternative. Datasheet C39173, 11/2011.
Eberline AC-3-7 / AC-3-8 Now Thermo Fisher Scientific
alpha scintillation probe
- Detector
- Alpha scintillators (CP-1 connector)
- Detects
- alpha
Hand probes for alpha surface contamination.
Probe for the ASP-2 and ASP-2e ratemeters (up to 1500 V).
Eberline ASP-2 Now Thermo Fisher Scientific
microprocessor portable ratemeter for external probes
- Detector
- None built in; accepts GM, gas proportional and scintillation probes (MHV, BNC or CP-1 connectors)
- Detects
- depends on probe (alpha, beta, gamma, neutron)
- Range
- Probe HV 300 to 2500 V; thresholds 0.5 to 60 mV; units CPS, CPM, DPS, DPM, Bq, R/hr, Sv/hr, Gy/hr, Rem/hr
General-purpose portable ratemeter that reads out contamination or dose rate from the probe plugged into it.
Basic model: ratemeter mode only. Set up with the WinASP2 host program; 3 C cells, about 100 h.
Eberline ASP-2e Now Thermo Fisher Scientific
enhanced microprocessor portable ratemeter for external probes
- Detector
- None built in; accepts GM, gas proportional and scintillation probes (MHV, BNC or CP-1 connectors)
- Detects
- depends on probe (alpha, beta, gamma, neutron)
- Range
- Probe HV 300 to 2500 V; thresholds 0.5 to 60 mV; units CPS, CPM, DPS, DPM, Bq, R/hr, Sv/hr, Gy/hr, Rem/hr
General-purpose portable ratemeter that reads out contamination or dose rate from the probe plugged into it.
Adds integrated dose, scaler counts and a pulse height window for energy-proportional probes (same manual as ASP-2).
Eberline E-140 Now Thermo Fisher Scientific
GM count rate meter (frisker meter for pancake GM probes)
- Detector
- No internal detector; takes an external GM probe on a BNC connector (at Hanford used with the pancake GM probe, P-11 type). Adjustable high voltage 200 to 1500 V.
- Detects
- Set by the probe; with a pancake GM probe, beta and gamma contamination.
- Range
- Meter 0 to 1,000 cpm with X1, X10, X100 ranges, upper limit 100,000 cpm (some units have a 0 to 500 cpm dial); linearity ±5% of full scale; response adjustable 2 to 10 seconds
Count rate meter that techs pair with a pancake GM probe for frisking and contamination surveys.
E-140B, BNW and BNW-1 are essentially identical. Attached speaker with its own on/off switch; phone jack for headphones or a scaler; BAT TEST position; RESET button; FAST/SLOW response switch. 2 D cells, about 200 hours. 1.5 kg (3.4 lb); 22 x 9.8 x 18 cm (8.5 x 3.9 x 6.9 in) with handle and probe clip.
Eberline E-600 Now Thermo Fisher Scientific
digital multi-probe portable radiation monitor (ratemeter, smart probe)
- Detector
- None internal. Takes Eberline smart probes (calibration and setup stored in the probe handle) or conventional probes, covering GM, scintillation, proportional, ion chamber and neutron detectors. Probe connector can carry over 2,000 V. Common Hanford pairings: SHP-380AB, SHP-380A (100 cm2 alpha/beta and alpha scintillators) and the NRD neutron rem ball.
- Detects
- Set by the probe: alpha, beta, gamma, neutron or energy windows. Up to three channels with independent alarms; alpha, beta-gamma and total channels can be shown at the same time.
- Range
- Units selectable by probe setup: cps, cpm, dps, dpm, Bq, dpm/100 cm2, Sv/h, Gy/h, R/h, rem/h and integrated counts, Sv, Gy, R, rem. Full-scale range changes by decades with up/down keys or auto-ranging; the digital display stays accurate to twice the meter scale. Simultaneous analog bar graph and digital readout. Modes: ratemeter, integrate, scaler, peak trap, background, check.
General purpose digital survey meter that techs swap between alpha/beta contamination, gamma and neutron probes without recalibrating.
Technical Manual rev. July 1995 (the E-600 was designed to replace a wide range of older portables). 22.9 x 10.5 x 15.3 cm (9.0 x 4.1 x 6.0 in); 1.53 kg (3.4 lb) with batteries. 3 alkaline C cells, typically 50 to 100 hours. Operating -20 to +50 C, storage -30 to +70 C, 0 to 95% non-condensing humidity. Log memory for 500 results. Windows interface program for setup and calibration. Hanford notes: audible output limited to 75 cps (4500 cpm); slow/medium/fast smart-channel response 20/10/3 s. Alpha check source is often a 0.85 nCi (about 2000 dpm) Th-230 source on the side of the case.
Eberline ESP-1 Now Thermo Fisher Scientific
"Smart Portable"
microcomputer-based digital portable survey meter (smart portable)
- Detector
- None internal. Most Eberline GM, proportional or scintillation detectors for alpha, beta, gamma or neutron can be used, on an MHV-series connector. Detector bias 350 to 2300 Vdc, set by the calibrator for the detector in use.
- Detects
- Set by the probe: alpha, beta, gamma or neutron.
- Range
- Basic unit is counts per second. Modes: Rate Meter (slow and fast response that switch automatically with count rate), Scaler, Inquiry/Calibration. Two-line, 16-character alphanumeric LCD with a bar graph of 1 in 48 resolution.
Digital count rate and scaler instrument for contamination and radiation surveys with a choice of Eberline probes.
From the ESP-1 Technical Manual (February 1992). 26.0 x 12.7 x 12.7 cm (10.25 x 5.0 x 5.0 in), about 1.72 kg (3.8 lb) with batteries, without probe. Six C cells, about 250 hours on carbon-zinc cells (no display light). -20 to +50 C. Seven front-panel buttons (ON/OFF, MODE, RESET, LIGHT, +, -, SPKR); 2000 Hz audible alarm.
Eberline HP-100 series (HP-100C, HP-100CGS) Now Thermo Fisher Scientific
gas flow proportional contamination probe
- Detector
- 100 cm2 gas flow proportional (MHV connector)
- Detects
- alpha, beta
Hand probe for alpha and beta surface contamination.
Probe for the ASP-2 and ASP-2e ratemeters (up to 1900 V).
Eberline HP-190A Now Thermo Fisher Scientific
end window GM probe
- Detector
- End window GM detector (BNC)
- Detects
- beta
Hand probe for beta surface contamination.
Probe for the ASP-2 and ASP-2e ratemeters (up to 900 V).
Eberline HP-210 Now Thermo Fisher Scientific
"pancake" Current
pancake GM probe
- Detector
- pancake Geiger-Mueller detector; thin mica window 1.4 to 2.0 mg/cm2 protected by a stainless steel screen; power supply 900 +/- 50 V
- Detects
- alpha and beta
Hand probe for daily contamination surveys of personnel, tables, floors and equipment.
Useful beta sensitivity down to 40 keV, alpha sensitive above 3 MeV; HP-210T adds tungsten shielding for beta in gamma fields.
Eberline HP-220A Now Thermo Fisher Scientific
shielded/directional GM probe
- Detector
- Shielded/directional GM detector (BNC)
- Detects
- gamma
Shielded, directional gamma probe.
Probe for the ASP-2 and ASP-2e ratemeters (up to 600 V).
Eberline HP-260 Now Thermo Fisher Scientific
"pancake"
pancake GM probe
- Detector
- pancake GM tube with thin mica window 1.4 to 2.0 mg/cm2, screen 79% open; active area 15.5 cm2; operating voltage 900 +/- 50 Vdc
- Detects
- alpha, beta, gamma
Hand probe for contamination surveys; background sensitivity about 3600 cpm per mR/hr (Cs-137).
Same detector specification as the HP-360, which differs only in its housing.
Eberline HP-330 Now Thermo Fisher Scientific
sealed gas proportional probe
- Detector
- Sealed gas proportional detector (MHV connector)
- Detects
- alpha, beta
Hand probe for alpha and beta surface contamination without a gas bottle.
Probe for the ASP-2 and ASP-2e ratemeters (up to 1800 V).
Eberline HP-340 Now Thermo Fisher Scientific
alpha/beta scintillation probe
- Detector
- Alpha/Beta scintillator (MHV connector)
- Detects
- alpha, beta
Hand probe for alpha and beta surface contamination.
Probe for the ASP-2 and ASP-2e ratemeters (up to 1000 V).
Eberline HP-350 Now Thermo Fisher Scientific
alpha scintillation probe
- Detector
- Alpha scintillator (MHV connector)
- Detects
- alpha
Hand probe for alpha surface contamination.
Probe for the ASP-2 and ASP-2e ratemeters (up to 1500 V).
Eberline HP-360 Now Thermo Fisher Scientific
pancake GM probe
- Detector
- Pancake GM detector (BNC)
- Detects
- beta
Pancake GM hand probe for contamination surveys.
Probe for the ASP-2 and ASP-2e ratemeters (up to 900 V).
Eberline HP-370 Now Thermo Fisher Scientific
pancake low energy gamma probe
- Detector
- Pancake low energy gamma detector (MHV connector)
- Detects
- low energy gamma
Hand probe for low-energy gamma emitters.
Probe for the ASP-2 and ASP-2e ratemeters.
Eberline HP-380A Now Thermo Fisher Scientific
alpha scintillation probe
- Detector
- 100 cm2 alpha scintillator (MHV connector)
- Detects
- alpha
100 cm2 hand probe for alpha surface contamination.
Probe for the ASP-2 and ASP-2e ratemeters.
Eberline HP-380AB Now Thermo Fisher Scientific
alpha/beta scintillation probe
- Detector
- 100 cm2 alpha/beta scintillator (MHV connector)
- Detects
- alpha, beta
100 cm2 hand probe for alpha and beta surface contamination.
Probe for the ASP-2 and ASP-2e ratemeters (up to 1000 V).
Eberline LEG-1 Now Thermo Fisher Scientific
low energy gamma scintillation probe
- Detector
- Low energy gamma scintillator, 1 in. dia x 0.04 in. thick NaI crystal (CP-1 connector)
- Detects
- low energy gamma
Probe for low-energy gamma and X-rays.
Probe for the ASP-2 and ASP-2e ratemeters (up to 1500 V).
Eberline PG-2 Now Thermo Fisher Scientific
low energy gamma/X-ray scintillation probe
- Detector
- Low energy gamma/X-ray scintillator, 2 in. dia x 0.08 in. thick NaI crystal (CP-1 connector)
- Detects
- gamma, X-ray (low energy)
Probe for low-energy gamma and X-ray emitters.
Probe for the ASP-2 and ASP-2e ratemeters (up to 1300 V).
Eberline RM-14 Now Thermo Fisher Scientific
GM count rate meter
- Detector
- None built in; BNC detector input with 900 V high voltage, made primarily for a Geiger-Mueller (G-M) detector (scintillation detectors with modification)
- Detects
- as the attached detector; primarily G-M
- Range
- 0-500 cpm meter scale with X1, X10, X100 multipliers (500; 5000; 50,000 cpm full scale)
Read out a G-M probe as a count rate with audible clicks and an adjustable locking alarm.
AC line or internal 6 V gel battery (about 50 hours); fast/slow response switch; recorder and scaler outputs.
Eberline RM-20 Now Thermo Fisher Scientific
alarming count rate meter (ratemeter)
- Detector
- None built in; charge-sensitive input and variable high voltage (0 to 2.5 kV scale) for Geiger-Mueller, scintillation and proportional detectors
- Detects
- as the attached detector (GM, scintillation, proportional)
- Range
- RM-20-1: 0-500, 0-5k, 0-50k, 0-500k cpm; RM-20-5: 0-10, 0-100, 0-1k, 0-10k cps
Read out a contamination probe as a count rate with audible clicks and an adjustable locking alarm.
RM-20 series units are identical except meter face; AC powered, optional internal 12 V battery; RM-25 later replaced it.
Eberline RM-25 Now Thermo Fisher Scientific
microcomputer-based count rate meter and frisker station
- Detector
- None built in; accepts Eberline GM, gas proportional and scintillation detectors, 500 to 2500 VDC, MHV connector
- Detects
- as the attached detector (GM, gas proportional, scintillator/PM tube)
- Range
- Analog meter marked 0 to 500 linearly; range multipliers x1, x10, x100, x1K
Frisker station: reads a contamination probe as a count rate with audible counts and alarm.
Enhanced RM-14/RM-20 replacement; setup and calibration by PC (WinRM25); 110-130 VAC or 12-14 VDC.
Eberline SPA-1A Now Thermo Fisher Scientific
alpha scintillation probe for swipes
- Detector
- Alpha scintillator, 1 in. diameter zinc sulfide element (MHV connector)
- Detects
- alpha
Alpha scintillator for particulate filters and swipe samples.
Probe for the ASP-2 and ASP-2e ratemeters (up to 1400 V).
Eberline SPA-3 Now Thermo Fisher Scientific
gamma scintillation probe
- Detector
- High sensitivity gamma scintillator, 2 in. dia x 2 in. NaI crystal (CP-1 connector; SPA-3MHV has MHV)
- Detects
- gamma
High sensitivity gamma scintillation probe.
Probe for the ASP-2 and ASP-2e ratemeters (up to 1600 V).
Eberline SPA-6 Now Thermo Fisher Scientific
gamma scintillation probe
- Detector
- Medium sensitivity gamma scintillator, 2 in. dia x 2.4 in. plastic element (MHV connector)
- Detects
- gamma
Medium sensitivity gamma probe.
Probe for the ASP-2 and ASP-2e ratemeters (up to 1600 V).
Eberline SPA-8 Now Thermo Fisher Scientific
gamma scintillation probe
- Detector
- Gamma scintillator, 1 in. dia x 1 in. NaI crystal (MHV connector)
- Detects
- gamma
General gamma scintillation probe.
Probe for the ASP-2 and ASP-2e ratemeters (up to 1500 V).
Eberline SPA-9 Now Thermo Fisher Scientific
iodine gamma scintillation probe
- Detector
- Gamma scintillator for iodine, 2 in. dia x 0.5 in. NaI crystal (MHV connector)
- Detects
- gamma (iodine)
Gamma probe for radioiodine.
Probe for the ASP-2 and ASP-2e ratemeters (up to 1600 V).
Ludlum Model 3
Current
analog GM/scintillation count rate meter (ratemeter)
- Detector
- no built-in detector; used with external GM or scintillation detectors; high voltage adjustable 400 to 1500 Vdc
- Detects
- alpha, beta, gamma, depending on the probe attached
- Range
- typically 0-200 mR/hr, or 0-500,000 cpm (4 linear ranges, x0.1, x1, x10, x100)
General purpose analog survey meter; the probe plugged in decides what it reads.
Fast (4 s) or slow (22 s) meter response; two D cells, typically more than 2000 hours.
Ludlum Model 12
Current
analog count rate meter for GM, scintillation or proportional detectors
- Detector
- no built-in detector; external GM, scintillation or proportional detectors; high voltage adjustable 400 to 2500 volts
- Detects
- alpha, beta, gamma, depending on the probe attached
- Range
- typically 0-500,000 counts/minute (CPM); multipliers x1, x10, x100, x1000
Analog ratemeter for contamination and gamma surveys; wide high voltage range lets it run gas proportional probes like the 43-68.
Input sensitivity adjustable -1 to -100 mV; fast (4 s) or slow (22 s) response.
Ludlum Model 26
"frisker" Current
integrated GM pancake frisker (meter and probe in one housing)
- Detector
- pancake GM (Geiger-Mueller) detector, stainless steel screen (79% open); active window area 15.51 cm2, open 12.26 cm2
- Detects
- alpha, beta, gamma; energy range 16.3 keV to 1.3 MeV
- Range
- 0.1 cps to 1.99 kcps; 1 cpm to 99.9 kcpm
One-handed frisking of people and objects for contamination.
RATE, MAX and COUNT modes (count up to 20 min); two AA batteries, about 1000 hours; Model 26-3 adds mR/hr and dpm units.
Ludlum Model 43-1
Current
ZnS alpha scintillation detector
- Detector
- ZnS(Ag) scintillator on a 3.8 cm (1.5 in.) magnetically shielded photomultiplier tube; window 1.2 mg/cm2 metalized polyester (1.3 mg/cm2 recommended for outdoor use); 83 cm2 active, 75 cm2 open area; operating voltage typically 500 to 1200 V
- Detects
- alpha only
- Range
- efficiency 33% for 239Pu (4 pi); background 3 cpm or less; uniformity within 10% of average reading
Large-area alpha probe for surface contamination surveys and smear counting with a ratemeter or scaler.
Part number 47-1516. Series C connector; 12.2 x 24.9 cm (Dia x L); 0.9 kg (2 lb). The Model 180-16 wide diameter sample holder (up to 10.2 cm / 4 in. samples) is an option. The 43-1-1 variant is an alpha-beta detector covered by the same manual (Jan 2010, combined 43-1 and 43-1-1).
Ludlum Model 43-5
Current
ZnS alpha scintillation probe
- Detector
- ZnS(Ag) scintillator, window typically 0.8 mg/cm2 metalized polyester; active area 79 cm2 (open 52 cm2); 38 mm photomultiplier tube
- Detects
- alpha
Alpha-only survey probe for surfaces; efficiency 13% (4pi) for Pu-239, background 3 cpm or less.
Operating voltage typically 500 to 1200 V; used with general purpose ratemeters and scalers.
Ludlum Model 43-37 (and 43-37-1)
"floor monitor probe" Current
large-area gas proportional floor detector
- Detector
- gas proportional, P-10 (10% methane, 90% argon) recommended; 43-37: 584 cm2 (90.5 in2) active, 434 cm2 open, 2.5 x 15.9 x 46.4 cm; 43-37-1: 821 cm2 (127 in2) active, 609 cm2 open, 2.5 x 15.9 x 64.1 cm; windows of 0.8 mg/cm2 aluminized Mylar (0.4, 3.9 or 7.9 mg/cm2 available on the 239-1F); operating voltage alpha typically 1000 to 1200 V (43-37-1: 1100 to 1400 V), beta-gamma typically 1600 to 1800 V (43-37-1: 1700 to 1800 V)
- Detects
- alpha and beta (gamma under 1%)
- Range
- 43-37-1 typical efficiency: gross counting 20% 239Pu, 30% 99Tc, 30% 90Sr/90Y; alpha-beta counting 17.5% 239Pu, 20% 99Tc, 20% 90Sr/90Y; background alpha under 10 cpm, beta-gamma 1000 to 1800 cpm in a 10 µR/hr field; threshold typically 2 to 4 mV (43-37-1: 4 mV)
Large-area gas proportional probe that rides on the 239-1F cart to scan floors for alpha and beta contamination.
43-37 figures are from the Model 239-1F data sheet (efficiency 239Pu 17%, 90Sr/90Y 25%, gamma under 1%); 43-37-1 figures are from its own data sheet (part number 47-2243). The 239-1F manual covers the detector. 43-37-1: 2.4 kg (5.4 lb). Needs a gas bottle, regulator and counting instrument.
Ludlum Model 43-68
"gas proportional probe" Current
gas proportional alpha-beta probe
- Detector
- rechargeable gas proportional detector, P-10 counting gas; window typically 0.8 mg/cm2 metalized polyester; active area 126 cm2 (open 100 cm2)
- Detects
- alpha and beta (less than 1% gamma)
Large-area alpha/beta contamination probe run on P-10 gas with a ratemeter or the Model 2360.
4pi efficiency Pu-239 20%, Sr-90/Y-90 20%; alpha plateau typically 1100-1400 V, beta-gamma 1600-1800 V.
Ludlum Model 43-89
Current
ZnS/plastic alpha-beta scintillation probe
- Detector
- ZnS(Ag) adhered to 0.25 mm plastic scintillator; window 1.2 mg/cm2 metallized polyester; active area approx. 125 cm2 (open approx. 100 cm2)
- Detects
- simultaneous alpha and beta
Large-area alpha and beta contamination probe for floors, tools and equipment.
4pi efficiency Pu-239 17.5%, Tc-99 15%, Sr-90/Y-90 17.5%; suggested instruments Model 2360, 3002, 3003.
Ludlum Model 43-93
Current
ZnS/plastic alpha-beta scintillation probe
- Detector
- ZnS(Ag) adhered to 0.25 mm plastic scintillator; window 1.2 mg/cm2 metalized polyester; active area 100 cm2 (open 88 cm2)
- Detects
- simultaneous alpha and beta
Alpha and beta contamination probe read separately on a dual-channel meter such as the 2360.
4pi efficiency Tc-99 15%, Sr-90/Y-90 20%, Pu-239 20%; background alpha 3 cpm or less, beta 300 cpm or less.
Ludlum Model 44-2
Current
NaI gamma scintillation probe
- Detector
- scintillator, 2.5 x 2.5 cm (1 x 1 in.) (diameter x thickness); 2.9 cm photomultiplier tube
- Detects
- gamma; recommended energy range 20 keV to 1.5 MeV
Sensitive gamma probe for finding sources and low-level gamma surveys; about 175 cpm per µR/hr (Cs-137).
Background 1800 cpm in a 10 µR/hr field; energy dependent, so not for dose measurements without correction.
Ludlum Model 44-3
Current
low energy gamma NaI scintillation detector
- Detector
- 2.5 cm (1 in.) diameter x 1 mm thick NaI(Tl) crystal, entry window 18.4 mg/cm2, 5 cm2 (0.8 in2) active and open area, on a 3.8 cm (1.5 in.) photomultiplier tube; operating voltage typically 500 to 1200 V
- Detects
- 125I and other low energy gamma; recommended energy range 10 to 60 keV; energy dependent
- Range
- sensitivity 675 cpm per µR/hr (125I); background under 250 cpm at 10 µR/hr; efficiency 33.5% for 125I (based on 129I efficiency of 18%)
Thin crystal probe for 125I and other low energy gamma surveys, such as thyroid checks and contamination in iodine work.
Part number 47-1533. Series C connector; -20 to 50 C; 5.1 x 17.8 cm (Dia x L); 0.5 kg (1 lb). Model 180-1, 180-1L and 180-24 sample holders are options.
Ludlum Model 44-7
Current
end window GM detector
- Detector
- end window halogen-quenched GM; 2.0 +/- 0.3 mg/cm2 mica window; 6 cm2 (0.93 in2) active, 5 cm2 (0.78 in2) open; operating voltage 900 V
- Detects
- alpha, beta, gamma; energy dependent
- Range
- efficiency (4 pi) 14C 2%, 90Sr/90Y 10%, 99Tc 6%, 239Pu 7%, 125I 0.1%; sensitivity about 2100 cpm per mR/hr (137Cs); background 40 cpm; dead time about 200 microseconds
Small-window GM probe for alpha, beta and gamma contamination surveys and counting wipes or planchets.
Part number 47-1536. Series C connector; -15 to 50 C (certifiable -40 to 65 C); 4.6 x 14.7 cm (Dia x L); 0.5 kg (1 lb). Model 180-4 sample holder is an option.
Ludlum Model 44-9
"pancake" Current
pancake GM probe
- Detector
- pancake-type, halogen-quenched GM; mica window 2.0 +/- 0.3 mg/cm2; active window area 15.51 cm2 (open 12.26 cm2)
- Detects
- alpha, beta, gamma
Standard hand probe for frisking and contamination surveys; 3300 cpm per mR/hr (Cs-137 gamma).
Typical 4pi efficiency 22% Sr-90/Y-90, 5% C-14; operates at 900 Vdc; optional dose filter flattens energy response.
Ludlum Model 44-10
"2x2" Current
2 x 2 in. NaI gamma scintillation detector
- Detector
- 5.1 x 5.1 cm (2 x 2 in.) NaI scintillator on a 5.1 cm (2 in.) magnetically shielded photomultiplier tube; operating voltage 500 to 1200 V; IP65
- Detects
- gamma; recommended energy range 30 keV to 3.0 MeV (manual: high-energy gamma 60 keV to 2 MeV); energy dependent
- Range
- sensitivity typically 900 cpm per µR/hr (137Cs); background 9750 cpm in a 10 µR/hr field; efficiency 125I 4%, 57Co 20%, 137Cs 9%, 60Co 15%
Probe that makes a ratemeter or scaler a sensitive low-level gamma survey and sample counting instrument.
Part number 47-1540. Series C connector; -20 to 50 C (certifiable -40 to 65 C); 6.6 x 27.9 cm (Dia x L); 1.0 kg (2.3 lb). Model 180-7 and 180-9 sample holders (180-9 has 3.81 cm lead housing) are options.
Ludlum Model 177
Current
alarm ratemeter (benchtop frisker)
- Detector
- none built in; takes GM or scintillation detectors (high voltage adjustable 400 to 1500 V, input sensitivity adjustable -10 to -100 mV)
- Detects
- depends on the detector connected (GM or scintillation)
- Range
- 4 ranges (x1, x10, x100, x1k) on a 0 to 500 cpm linear analog meter dial, so 0 to 500,000 cpm full scale; linearity within 5% of full scale; adjustable alarm 0 to 150% of full scale (latching as shipped); response fast 2.2 s or slow 22 s to 90% of full scale
Line-powered benchtop ratemeter with an alarm that is set up as a frisking station with a pancake or scintillation probe.
Part number 48-1632. 6 V sealed lead-acid battery built in, about 50 h without alarm; 95 to 250 Vac wall adaptor; recorder, pulse and alarm outputs on a 9-pin D connector; 1.9 kg (4.2 lb); 12.7 x 20.3 x 15.2 cm (H x W x L). Other versions exist (177-50, 177-56, 177-61).
Ludlum Model 239-1F
"floor monitor" Current
gas proportional floor contamination monitor (detector on a cart)
- Detector
- Model 43-37 (584 cm2 active) or 43-37-1 (821 cm2 active) gas proportional detector on a roll-around cart; P-10 gas; window 0.8 mg/cm2 aluminized Mylar (0.4, 3.9 or 7.9 available); detector height adjustable 0.32 to 7.6 cm (0.125 to 3 in.); operating voltage alpha typically 1000 to 1200 V, beta-gamma 1600 to 1800 V; threshold typically 2 to 4 mV
- Detects
- alpha, beta (gamma under 1%); gas proportional, floor contamination
- Range
- efficiency (4 pi) 239Pu 17%, 90Sr/90Y 25%, gamma <1%; gas flow 0 to 100 cc/min, typically 50 cc/min; operates on static charge for 2 hours
Cart-mounted floor scanner used with a survey meter and gas bottle to find alpha and beta contamination on floors.
Needs a counting gas bottle, regulator and survey meter (data sheet lists the Models 3002, 3003, 12, 2350-1 and 2360 as configurations). Cart 106.7 x 40.6 x 69.9 cm (H x W x L) without detector; 11.4 kg (25 lb) without gas bottle and meter. The 2010 manual quotes older specs (e.g. 35 cc/min, 20 lb).
Ludlum Model 2224-1
Current
alpha-beta scaler ratemeter
- Detector
- none built in; takes proportional and dual phosphor scintillation detectors; high voltage adjustable 400 to 2000 V
- Detects
- simultaneous but separate alpha and beta counting with alpha/beta audio discrimination (front switch selects alpha, beta, or alpha plus beta)
- Range
- analog meter 0 to 500 cpm with x1, x10, x100, x1000 multipliers; 6-digit LCD scaler; count time 0.1, 0.5, 1 or 2 min; linearity within 10% of true value; response typically 2 to 11 s from 10% to 90%
Portable alpha-beta ratemeter and scaler used with a gas proportional or phoswich probe for contamination surveys and smear counting.
Part number 48-2679. 2 D cells, over 350 h with alkaline; overload indicator; 1.6 kg (3.5 lb); 16.5 x 8.9 x 21.6 cm (H x W x D) with handle.
Ludlum Model 2241-2
Current
digital survey meter with built-in scaler (ratemeter/scaler)
- Detector
- no built-in detector; stores settings for two external detectors (DET1/DET2 switch)
- Detects
- alpha, beta, gamma, depending on the probe attached
- Range
- auto-ranging 0.0 µR/hr to 9999 R/hr; 0.000 µSv/h to 9999 Sv/h; 0 C/m to 999 C/m; 0 C/s to 100 kC/s
Digital ratemeter and scaler; swap between a pancake GM and a gamma scintillator without recalibrating.
Ratemeter or scaler mode (1 to 9999 s); two D cells, typically 200 hours; 2241-2RK kit adds a 44-9 and 44-2.
Ludlum Model 2360
Current
dual-channel alpha/beta scaler/ratemeter data logger
- Detector
- no built-in detector; suggested detectors are dual phosphor scintillation and gas proportional probes
- Detects
- simultaneous alpha and beta
- Range
- meter dial 0-500 cpm; range multipliers x1, x10, x100, x1000; 6-digit scaler, count times 0.1 to 60 minutes
Counts alpha and beta together from one probe, with logged results for contamination surveys and smear counting.
Stores 550 data points; RS-232 port; six separate alarms; typical battery life 250 hours.
Ludlum Model 2401-P
Current
pocket GM survey meter with built-in pancake detector
- Detector
- integral pancake GM with stainless steel protective screen (79% open); window area 15.51 cm2 (2.4 in2) active, 12.26 cm2 (1.9 in2) open
- Detects
- alpha, beta, gamma; energy dependent (pancake over-responds to gamma between 20 and 150 keV; the 2401-P DOSE variant adds a sliding filter flat within 20% of 137Cs from 20 keV to 1.2 MeV)
- Range
- 3 ranges (x1, x10, x100); default dial 0 to 6.6 kcpm and 0 to 2 mR/hr (other meter faces available, e.g. 0 to 500 cpm, 0 to 10 cps / 0 to 1.9 µSv/h); sensitivity about 3300 cpm per mR/hr (137Cs); efficiency 14C 5%, 90Sr/90Y 22%, 99Tc 19%, 99mTc <1%, 32P 32%, 239Pu 15%; response about 11 s from 10% to 90%
Pocket-size pancake frisker and contamination survey meter for personnel and surface checks.
Part number 48-2875. No alarm. 9 V battery, about 250 h with alkaline; 4.6 x 8.4 x 13.5 cm; 0.4 kg (0.9 lb) with battery; click-per-event audio can be switched off (QUIET). The technical manual also covers the 2401-EC, 2401-EW and 2401-EC2.
Mirion RDS-80
Current
GM surface contamination meter
- Detector
- end window GM tube, 1.5 - 2 mg/cm2 mica window, active surface area 15.2 cm2
- Detects
- alpha >2 MeV, beta Emax >100 keV, gamma and X-rays 5 keV to 3 MeV
- Range
- surface contamination 1 to 100 000 cps or 0.01 to 1 000 000 Bq/cm2
Handheld pancake-style contamination meter that reads cps or Bq/cm2.
Histogram of 480 points; IrDA port with RDS-CSW software; 2 AA batteries, at least 2000 hours.
Thermo Scientific RadEye B20
Current
pancake GM multi-purpose survey meter (contamination and dose rate)
- Detector
- 1 pancake GM tube, window diameter 44 mm (1.7 in.), 1.8 - 2.0 mg/cm2
- Detects
- alpha, beta, gamma and X-ray
- Range
- contamination 0 - 10 kcps (B20), 0 - 500 kcps (B20-ER); dose rate 0.2 µSv/h - 2 mSv/h (B20), 0.2 µSv/h - 100 mSv/h (B20-ER)
Pocket frisker and survey meter; scaler, ratemeter and dose rate modes.
Energy range 17 keV - 3 MeV with optional filters; optional alpha blocker separates alpha from beta; 300 g, 2 AAA batteries.